CBCT-Based Synthetic CT Image Generation Using Conditional Denoising Diffusion Probabilistic Model

Junbo Peng, Richard L. J. Qiu, Jacob F Wynne, Chih-Wei Chang, Shaoyan Pan, Tonghe Wang, Justin Roper, Tian Liu, Pretesh R. Patel, David S. Yu, Xiaofeng Yang

arXiv:2303.02649v1physics.med-ph

TL;DR

CBCT artifacts and inaccurate HU values hinder quantitative ART applications. This paper develops a conditional DDPM to generate sCT from CBCT; the produced images show improved HU accuracy and fewer artifacts, with stated suitability for dose calculation.

  • Problem

    Artifacts and inaccurate HU values in CBCT hinder its use for quantitative applications such as ART.

    Source: §1 · p. 3
  • Method

    The paper develops a conditional DDPM framework to generate sCT from CBCT images.

    Source: §5 · p. 18
  • Results

    The produced sCT images showed improved HU accuracy and fewer artifacts than input CBCT, with significantly reduced metal artifacts reported.

    Source: §4 · p. 16, §5 · p. 18
  • Takeaways & Limitations

    The improved HU accuracy makes the produced sCT images suitable for dose calculation.

    Source: §5 · p. 18

    Artifact-reduction methods share a limitation: they may suppress specific artifact sources without correcting other artifact types.

    Source: §1 · p. 3

Abstract

from arXiv · show

Background: Daily or weekly cone-beam computed tomography (CBCT) scans are commonly used for accurate patient positioning during the image-guided radiotherapy (IGRT) process, making it an ideal option for adaptive radiotherapy (ART) replanning. However, the presence of severe artifacts and inaccurate Hounsfield unit (HU) values prevent its use for quantitative applications such as organ segmentation and dose calculation. To enable the clinical practice of online ART, it is crucial to obtain CBCT scans with a quality comparable to that of a CT scan. Purpose: This work aims to develop a conditional diffusion model to perform image translation from the CBCT to the CT domain for the image quality improvement of CBCT. Methods: The proposed method is a conditional denoising diffusion probabilistic model (DDPM) that utilizes a time-embedded U-net architecture with residual and attention blocks to gradually transform standard Gaussian noise to the target CT distribution conditioned on the CBCT. The model was trained on deformed planning CT (dpCT) and CBCT image pairs, and its feasibility was verified in brain patient study and head-and-neck (H&N) patient study. The performance of the proposed algorithm was evaluated using mean absolute error (MAE), peak signal-to-noise ratio (PSNR) and normalized cross-correlation (NCC) metrics on generated synthetic CT (sCT) samples. The proposed method was also compared to four other diffusion model-based sCT generation methods. Conclusions: The proposed conditional DDPM method can generate sCT from CBCT with accurate HU numbers and reduced artifacts, enabling accurate CBCT-based organ segmentation and dose calculation for online ART.

1. Introduction

CBCT artifacts and HU inaccuracies impede quantitative ART; this work proposes conditional DDPM translation from CBCT to sCT, with brain and H&N experiments reporting accurate HU values and reduced artifacts.

1. Introduction

  • CBCT is widely used in IGRT, but artifacts and inaccurate HU values hinder quantitative applications and CBCT-based ART.The cited artifacts include streaking, shading, cupping, and scatter contamination.Source: §1 · p. 3
  • Current approaches either correct CBCT artifacts or generate sCT images intended to approach pCT image quality.The introduction presents these as two major routes toward direct CBCT-based ART.Source: §1 · p. 3
  • Hardware modifications can reduce scatter artifacts but may lower system quantum efficiency and reconstructed-image signal-to-noise ratio.Examples include anti-scatter grids, primary-modulation beam filters, and lattice-shaped beam stoppers.Source: §1 · p. 3
  • Monte Carlo scatter estimation is accurate but computationally intensive, while analytical kernels are faster but less effective for heterogeneous structures; learning-based correction is artifact-source-specific.The source-specific limitation is described as common to artifact-reduction methods.Source: §1 · p. 3
  • Paired CBCT–dpCT translation can suffer anatomical mismatch; proposed alternatives include unpaired Cycle-GAN consistency and added structure-preserving modules or losses.Gradient-difference loss and attention mechanisms are cited as examples of added strategies.Source: §1 · pp. 3–4
  • DDPM and other diffusion models have been used for medical-image reconstruction, MRI–CT conversion, denoising, and generation.The introduction characterizes diffusion models as analytically principled, easy to train, and capable of state-of-the-art image quality.Source: §1 · p. 4
  • The proposed conditional DDPM trains on CBCT–dpCT pairs and gradually transforms Gaussian noise into an sCT conditioned on its CBCT counterpart.The dpCT serves as the target distribution and CBCT as the condition.Source: §1 · p. 4
  • Brain and H&N experiments found that conditional DDPM generated sCT with accurate HU values and reduced artifacts, supporting direct CBCT-based ART.The authors state that these results make direct CBCT-based ART implementation possible.Source: §1 · p. 4

2. Materials and Methods

This section describes paired CBCT–CT data preparation and a conditional DDPM for generating synthetic CT images, including alternative conditioning strategies and model implementation. It also outlines evaluation against CT ground truth using MAE, PSNR, and NCC.

2.1. Image acquisition and preprocessing

  • The study included 41 brain and 47 H&N patients; brain training used 4682 slices from 30 patients and testing used 500 slices from 11 others.For H&N, 4314 slices from 37 patients were used for training and 500 slices from 10 other patients for testing.Source: §2.1 · p. 4
  • CT images were deformably registered and resampled to match corresponding CBCT images, with identical data volume and voxel size; all images were cropped to 256 × 256 and normalized to [-1, 1] before network input.Merged body contours extracted from CBCT and CT images were applied to both images.Source: §2.1 · p. 4

2.2. Denoising diffusion probabilistic models (DDPM)

  • DDPM gradually corrupts target images with Gaussian noise, trains a network to predict that noise, and reverses the process from Gaussian noise to recover an image.The noise-prediction objective supports estimating the reverse transition at each step.Source: §2.2 · p. 5

2.3. The proposed conditional DDPM for synthetic CT from CBCT images

  • The proposed model concatenates the input CBCT with the noisy CT sample at each step, conditioning the noise estimator and reverse process on CBCT.This makes the task conditional image-to-image translation rather than generation of a random sample from the CT distribution.Source: §2.3 · p. 6

2.4. Compared conditional DDPM approaches

  • Uncond-1 applies ILVR during inference, while Uncond-2 refines unconditional sampling using a progressively weakened CBCT condition.ILVR uses a low-pass-filtered CBCT constraint to preserve high-level semantics; both approaches use an unconditionally trained model.Source: §2.4 · p. 8
  • Adap-Cond-1 and Adap-Cond-2 use adaptively noised CBCT conditions during training and sampling instead of the proposed method’s constant condition.Adap-Cond-1 conditions prediction on y_(t-1), whereas Adap-Cond-2 uses y_t so the CBCT and CT conditions have the same noise level.Source: §2.4 · p. 8

2.5. Implementation and evaluation

  • A single time-embedded U-net with attention modules and residual blocks predicts noise across 1000 steps, with variance scheduled linearly from β_1 = 10^-4.The same noise-prediction model is used at every diffusion step.Source: §2.5 · p. 11
  • MAE, PSNR, and NCC were calculated between sCT images and CT images taken as ground truth.Source: §2.5 · p. 11
  • MAE measures voxel-based HU difference between the original CT and sCT, while PSNR measures whether predicted sCT intensity is evenly or sparsely distributed.Source: §2.5 · pp. 11–12
  • NCC measures similarity between CT and sCT as a function of displacement.Source: §2.5 · p. 12

3. Results

In the brain and H&N studies, the proposed sCT reduced visible CBCT artifacts and improved HU fidelity. Quantitative metrics favored sCT over CBCT, while comparisons among diffusion-model schemes showed differing trade-offs in structural preservation and residual artifacts.

3.1. Visual quality improvement

  • Brain CBCT beam-hardening artifacts were significantly suppressed in sCT while fine structures, including contrast agent and maxillary sinus, were preserved.The paper notes that preserving these structures can improve organ-segmentation accuracy.Source: §3.1 · p. 12
  • In H&N images, sCT appeared artifact-free for two slices and suppressed artifacts better than dpCT in extremely metal-degraded slices.The CBCT images had severe streaking artifacts from motion and metal implants.Source: §3.1 · p. 13
  • Across 500 testing slices per study, sCT HU-histogram shapes and peaks were closer to dpCT than CBCT, indicating improved HU fidelity.Source: §3.1 · p. 13

3.2. Quantitative analysis

  • 25.99 HU MAE for brain sCT versus 40.63 HU for CBCT; H&N MAE fell to 32.56 HU from 38.99 HU.Brain sCT PSNR and NCC were 30.49 dB and 0.99, compared with 27.87 dB and 0.98 for CBCT; H&N PSNR and NCC improved slightly.Source: §3.2 · p. 14

3.3. Comparison studies

  • ILVR preserved a CBCT-like outline but showed distortions and artifacts, whereas Uncond-2 performed better and preserved more image detail.The text attributes Uncond-2's advantage to directly enforcing the image condition during sampling.Source: §3.3 · p. 15
  • Weak adaptive conditioning produced different failures: Adap-Cond-1 retained most anatomy with some distortion, while Adap-Cond-2 failed to preserve the body contour.Source: §3.3 · p. 15
  • Residual artifacts remained among the four compared methods, and weak guidance caused artifacts in some cases.Source: §3.3 · pp. 15–16

4. Discussion

The discussion presents conditional DDPM as a first reported diffusion approach to CBCT–CT translation and reports reduced artifacts in the H&N study. It also highlights limits involving reference mismatch, unsupervised conditioning, sampling time, and 2D modeling.

4. Discussion

  • To the authors’ knowledge, this was the first reported use of a diffusion model for CBCT–CT translation.Source: §4 · p. 16
  • Supervised medical image translation commonly lacks exactly matched training pairs, and matched CBCT–CT images are described as impossible to obtain.Source: §4 · p. 16
  • In the H&N study, conditional DDPM significantly reduced metal artifacts, producing sCT superior to the artifact-corrupted dpCT counterparts used for training.Source: §4 · pp. 16–17
  • Because dpCT is not anatomically matched to sCT, sCT–dpCT quantitative metrics are unfair to some extent, while sCT seems closer to artifact-free ground truth.Source: §4 · p. 17
  • The discussion proposes further study of unsupervised diffusion for sCT generation, noting that its conditional form requires unconditional training and conditional sampling.In this setup, standard conditional-probability condition enforcement is inaccessible, and possible solutions are left for future work.Source: §4 · p. 17
  • Stochastic Gaussian noise makes it difficult to incorporate prior information during training, leaving sampling quality dependent on network manifold learning and insufficient for inter-channel correlations such as DECT.Source: §4 · p. 17
  • Sampling time is a major obstacle to clinical use, although diffusion-model training time is comparable to one-step U-net models.With T steps, generation takes approximately T times as long as with U-net or GAN models, making online ART impractical.Source: §4 · p. 17
  • The method uses 2D slices without patch extraction; 3D patch-based models could better preserve structure and capture spatial relationships, while other diffusion alternatives may outperform DDPM.Source: §4 · p. 17

5. Conclusions

The conditional DDPM generates synthetic CT from CBCT, with improved HU accuracy and fewer artifacts than the input. These image improvements support dose calculation, online segmentation, and replanning.

  • The conditional DDPM generates synthetic CT (sCT) from CBCT images.Source: §5 · p. 18
  • Improved HU accuracy makes the generated sCT suitable for dose calculation.Source: §5 · p. 18
  • Fewer artifacts in sCT than in CBCT could improve CBCT-based online segmentation and replanning.Source: §5 · p. 18
  • The method supports further quantitative applications of CBCT and may enhance the potential for clinical CBCT-based adaptive radiotherapy.Source: §5 · p. 18