Physics-Informed Deep Learning for False Ventricular Tachycardia Alarm Reduction in the ICU

Athanasios Papastathopoulos-Katsaros, Alexandra Stavrianidi, Zhandong Liu

arXiv:2609.08992v2cs.LGeess.SP

TL;DR

False VT alarms are hard to distinguish from true arrhythmias because ICU artifacts can resemble VT. The paper combines a 1D SE-ResNet with auxiliary physics-informed ABP and data-driven PLETH reconstruction; on VTaC, it reports a ∼5-point Challenge Score gain over prior state-of-the-art within a 10 s real-time window.

  • Problem

    ICU artifacts can resemble true arrhythmias, making false VT alarms difficult to distinguish and contributing to alarm fatigue.

    Source: §1 · p. 1
  • Method

    A 1D SE-ResNet jointly classifies alarms and reconstructs ABP through a Windkessel simulation and PLETH through a data-driven decoder.

    Source: §1 · p. 1, §2.2 · p. 1, §2.4 · p. 2
  • Results

    A ∼5-point Challenge Score improvement over prior state-of-the-art was reported on VTaC using a strict 10 s real-time window without external pretraining data.

    Source: §1 · p. 1
  • Takeaways & Limitations

    The authors conclude that the physics-informed constraint drives performance gains while improving label efficiency and interpretability for clinical classification with limited labeled data.

    Source: §5 · p. 4

    Evaluation is limited to a single benchmark, ABP is available in only 36% of recordings, and the Windkessel approximation is lumped-parameter.

    Source: §4 · p. 4

Abstract

from arXiv · show

False ventricular tachycardia (VT) alarms are a leading contributor to alarm fatigue in intensive care units. We propose a deep learning framework combining a 1D SE-ResNet with ICU-realistic data augmentations and a physics-informed auxiliary reconstruction task based on the three-element Windkessel hemodynamic model, implemented as a differentiable forward simulation. By requiring the network's latent representation to produce physiologically plausible arterial pressure waveforms, artifact-driven ECG patterns are penalized while true VT remains coherent across modalities. Evaluated on the VTaC benchmark under a strict real-time protocol (10-second pre-alarm window), our method achieves a 5-point Challenge Score improvement over prior state-of-the-art. Ablation studies confirm that the physics-informed objective is the primary performance driver, providing gains in accuracy, 2x label efficiency, and more localized and clinically meaningful ECG segments.

1. Introduction

False VT alarms are difficult to distinguish from true arrhythmias because ICU artifacts can mimic VT, while high-sensitivity monitoring makes false positives common. The paper embeds physiological structure into learning and reports an improved VTaC Challenge Score under a 10-second real-time protocol.

1. Introduction

  • Sensor detachment, patient movement, and electrical interference can produce ECG artifacts that closely resemble true arrhythmias, confounding data-driven classifiers.VTaC provides over 5,000 multi-institutional ICU recordings with ECG, PLETH, and ABP for evaluation.Source: §1 · p. 1
  • Prior real-time methods achieved Challenge Scores up to 80.08, while higher-AUC foundation-model and retrospective approaches rely on external pretraining or post-alarm information.These approaches remain vulnerable to severe sensor noise and previously unseen artifact patterns.Source: §1 · p. 1
  • 85.08 ± 1.65 Challenge Score, a ∼5-point improvement over prior state-of-the-art, using a strict 10 s real-time window without external pretraining data.The method jointly optimizes classification with physics-informed ABP reconstruction and data-driven PLETH reconstruction.Source: §1 · p. 1

2. Methods

The methods combine multi-channel ICU data and a 1D SE-ResNet with realistic augmentations, class-imbalance handling, and auxiliary waveform reconstruction. A differentiable three-element Windkessel simulation supplies structured regularization alongside data-driven PLETH reconstruction.

2.1. Dataset and Preprocessing

  • 5,037 expert-annotated VTaC alarm events from three U.S. hospitals are split patient-wise 80-10-10, using only the final 10 s before each alarm.Fourteen canonical channels are mapped to fixed slots with zero-filled unavailable channels and a concatenated availability mask; each channel is independently z-normalized.Source: §2.1 · p. 1

2.2. Architecture

  • The 1D SE-ResNet1D backbone uses a three-branch multi-scale stem and four feature stages, with shared features feeding classification and ABP and PLETH reconstruction heads.Models contain 2–9M trainable parameters depending on auxiliary heads; hyperparameters and the decision threshold are selected using validation data.Source: §2.2 · pp. 1–2, Figure 1 · p. 2

2.3. Augmentation and Class Imbalance

  • Five ICU-oriented augmentations target signal degradation, while Asymmetric Focal Loss addresses class imbalance using a tunable false-negative weight.Augmentations include temporal jitter, Gaussian noise, channel dropout, amplitude scaling, and 0.1–0.5 Hz baseline wander.Source: §2.3 · p. 2

2.4. Physics-Informed Regularization

  • A six-parameter projection head drives a differentiable two-stage Windkessel simulation to reconstruct ABP as an auxiliary task, not as the model’s prediction target.A synthetic ejection proxy feeds the three-element model, whose exact exponential integration is stable for positive Rp and C.Source: §2.4 · p. 2
  • The ABP reconstruction loss regularizes shared features because artifact-driven representations are less compatible with coherent hemodynamics and incur higher reconstruction error.PLETH is reconstructed by a data-driven MLP; its availability is about 91% of recordings versus about 36% for ABP, and reconstruction losses are masked when channels are unavailable.Source: §2.4 · pp. 2–3

3. Results

The results compare model variants and prior work on VTaC using a Challenge Score that weights missed true alarms fivefold. Ablations report gains from physics-informed reconstruction, label efficiency, modality robustness, and more localized saliency.

3. Results

  • The primary Challenge Score is (TP + TN)/(TP + TN + FP + 5 · FN), weighting missed true alarms fivefold; reported scores are mean ± SD over five random seeds.Source: §3 · p. 3

3.1. Main Results

  • Table 1 compares model variants with prior work on VTaC using Challenge Score, reported as mean ± SD over five seeds where available.The table caption identifies augmentation, data-driven reconstruction, and physics-informed reconstruction variants.Source: Table 1 · p. 3, §3.1 · p. 3

3.2. Ablation Studies

  • Without augmentations, physics-informed reconstruction improves the baseline by +1.42 Challenge Score points versus +0.57 for data-driven reconstruction; with augmentations, it improves over augmentation alone by +3.37 versus +2.45.Source: §3.2 · p. 3
  • At 5% labels, the physics-informed model scores 64.74±2.94, nearly matching the augmentation-only baseline at 10% labels (65.04 ± 1.98), and is the only multi-task variant to beat baseline at every fraction.Table 2 evaluates label fractions of 5%, 10%, 20%, and 50%.Source: §3.2 · p. 3, Table 2 · p. 3
  • Sensor dropout reduces Challenge Score by −1.33 for ABP and −4.85 for PLETH; replacing signals with uncorrelated Gaussian noise reduces it by −4.98 for ABP and −6.42 for PLETH.Corrupted signals are reported as more damaging than missing signals; Table 2’s modality-robustness results are from a single run.Source: §3.2 · p. 3, Table 2 · p. 3
  • On 128 true-alarm samples, the physics-informed model produces more localized Grad-CAM saliency in 89.1% of cases, with mean Gini coefficient rising from 0.52 ± 0.11 to 0.65 ± 0.09.Figure 2 illustrates a false alarm that the baseline misclassifies but the physics-informed model correctly rejects.Source: §3.2 · p. 3, Figure 2 · p. 3

4. Discussion

False VT ECG artifacts can resemble true arrhythmias, while the Windkessel reconstruction penalizes artifact-driven representations that lack a coherent hemodynamic correlate. The model’s reconstruction accuracy is limited by the coarse approximation and ABP availability, and evaluation covers one benchmark.

  • Sensor detachment, movement, and electrical interference can create wide-complex, high-rate ECG patterns indistinguishable from VT on a single lead.Source: §4 · p. 3
  • The Windkessel head penalizes artifact-driven features because electrical artifacts lack a hemodynamic correlate, making simulated pressure incoherent and increasing reconstruction loss.Source: §4 · p. 4
  • ABP reconstruction loss decreases only approximately 2—3%, and the Windkessel model coarsely approximates measured waveforms; limitations also include ABP in 36% of recordings and evaluation on one benchmark.Despite modest reconstruction accuracy, the fixed dynamical structure propagates corrective gradients, unlike a flexible MLP decoder that could absorb artifacts.Source: §4 · p. 4

5. Conclusions

The conclusion presents a hemodynamic forward model as an auxiliary reconstruction task that provides an inductive bias for reducing false VT alarms.

  • Embedding a simple hemodynamic forward model as an auxiliary reconstruction task provides a powerful inductive bias for false VT alarm reduction.Source: §5 · p. 4