3D Computational Modeling of the Vascular Heat-Sink Effect in Microwave Ablation of Tumors Close to the Aorta

Youjin Kim, Kyungho Yoon, Minwoo Shin

arXiv:2609.12407v1math.NAcs.NA

TL;DR

Blood flow near the aorta can cool adjacent tumor tissue and increase the risk of incomplete microwave ablation, while patient-specific conditions are difficult to predict from imaging alone. The study uses coupled thermal-fluid simulations with mapped microwave heating to compare vessel distances and flow conditions. It reports the strongest reduction in tumor coverage at direct contact, a smaller reduction at 4 mm, and nearly unchanged coverage at 8.5 mm.

  • Problem

    Blood flow can suppress temperature rise near a large vessel and risk incomplete ablation, while patient-specific thermal behavior is difficult to predict from real-time imaging alone.

    Source: §1 · p. 2
  • Method

    The study uses coupled thermal-fluid simulation with an equivalent microwave heat source to compare tumor–aorta distances and blood-flow conditions.

    Source: §1 · p. 2, §2.4 · p. 5
  • Results

    Shorter tumor-to-aorta distance and the presence of the blood domain reduced interface temperature and tumor coverage, most clearly at 0 mm, less at 4 mm, and nearly not at 8.5 mm.

    Source: §5 · p. 8
  • Takeaways & Limitations

    Treatment planning for tumors near large vessels should consider tumor coverage, the vessel-side 60°C margin, and patient-specific vessel position alongside input power and central temperature.

    Source: §4.2 · p. 8

Abstract

from arXiv · show

Microwave ablation (MWA) is a percutaneous ablation that induces thermal necrosis by generating localized high-temperature regions within tumor tissue. However, when a tumor is located close to a large vessel such as the aorta, blood flow continuously removes heat from the ablation zone. This phenomenon causes the vascular heat-sink effect and thereby increases the risk of incomplete ablation. This study constructs an ANSYS-based numerical framework to investigate the heat-sink effect during MWA of tumors located close to the aorta. The computational model consists of the tumor, surrounding tissue, vessel wall, blood flow domain, and microwave antenna. Microwave-induced heating was represented by an equivalent heat source mapped from the High-Frequency Structure Simulator (HFSS)-derived volume loss density, rather than by direct electromagnetic-thermal coupling within the thermal-fluid solver. Two parametric studies were performed to quantify the heat-sink effect. First, the tumor-aorta distance was varied at D = 0 mm, 4 mm, and 8.5 mm to assess how vessel proximity affects temperature elevation and ablation zone formation. Second, at each tumor-aorta distance, three blood-flow conditions were compared: no blood, stagnant blood V = 0 m/s, and normal aortic blood flow V=0.65 m/s. This comparison enabled the heat-sink effect of blood flow to be independently evaluated. The results highlight the clinical need for simulation-based treatment planning in patients with tumors located close to major vessels.

1 Introduction

Microwave ablation heats tumor tissue to cause irreversible injury, but blood flow near a large vessel can cool the tumor edge and risk incomplete ablation. The study motivates computational treatment planning and examines how tumor–aorta distance and blood flow affect temperature and ablation.

1 Introduction

  • Microwave ablation uses microwave energy to create localized high-temperature regions that cause irreversible tumor-cell injury.The paper describes treatment temperatures of 60°C or higher as generally cytotoxic.Source: §1 · pp. 1–2
  • Blood flow near a large vessel removes heat from adjacent tissue, suppressing temperature rise and risking incomplete tumor ablation.Effective thermal ablation requires the entire tumor and a sufficient safety margin to reach the target temperature.Source: §1 · p. 2
  • Patient variation in tumor geometry and blood-flow conditions makes thermal behavior difficult to predict from real-time imaging alone.Coupled thermal-fluid simulation can quantify tumor coverage, vessel-side cooling, and the resulting ablation extent before treatment.Source: §1 · p. 2
  • The study compares tumor–aorta distances of 0, 4, and 8.5 mm under no-blood, stagnant-blood, and normal-flow conditions.It compares temperature contours and ablation-zone volumes to assess incomplete-ablation risk near large vessels.Source: §1 · p. 2

2 Method

The model combines tumor, surrounding tissue, aorta, blood, and antenna domains, then applies mapped microwave heating in a coupled thermal-fluid simulation. It varies vessel distance and blood-flow conditions to examine heat transfer and ablation-zone formation.

2.1 Geometry Construction

  • The computational geometry includes tumor, surrounding tissue, aorta, blood domain, and microwave antenna, with tumor–aorta distances of 0, 4, and 8.5 mm.The tumor geometry was derived from a 3D liver segmentation dataset, and the aorta from a vascular model scaled for the computational domain.Source: §2.1 · pp. 2–4

2.2 Material Properties

  • Biological properties were selected from prior MWA simulations and reported tissue ranges, while antenna-component thermal properties came from the ANSYS material database.Copper, stainless steel, and Teflon were assigned to antenna components.Source: §2.2 · pp. 4–5

2.3 HFSS-to-Fluent Heat Source Mapping

  • HFSS-derived volume loss density at 2.45 GHz and 20 W was mapped to Fluent as a nonuniform heat source near the antenna slot.A temperature-dependent attenuation factor reduced the source above 333.15 K; the electromagnetic field was not recalculated during transient thermal-fluid analysis.Source: §2.3 · p. 5

2.4 Physical Model

  • The coupled model calculates conduction in solid regions and convective heat removal in the blood domain, using SST k-ω for blood-flow turbulence.The simulations compare tumor temperature and ablation-zone formation under aortic cooling.Source: §2.4 · p. 5

3 Experiments

The experiments vary tumor–aorta distance and blood-flow condition to compare temperature reduction near the vessel and ablation-zone volume. The flow study distinguishes no blood, stagnant blood, and normal aortic flow.

3.1 Tumor-to-Aorta Distance Study

  • Tumor–aorta distance was set to 0, 4, or 8.5 mm, from direct contact to progressively larger tissue gaps.The comparison assesses vessel-side temperature reduction, ablation volume, and asymmetry of the 60°C region.Source: §3.1 · pp. 5–6
  • Across both study dimensions, the comparisons assess tumor temperature distribution, vessel-side cooling, and ablation-zone volume.Table 2 is identified as the quantitative comparison of ablation volumes across distances and blood-flow conditions.Source: §3.1 · p. 6, §3.2 · p. 6, Table 2 · p. 6

3.2 Blood Flow Condition Study

  • Blood-flow conditions were no blood, a blood domain at 0 m/s, and normal aortic flow at 0.65 m/s.The conditions isolate the effect of blood-domain presence and increasing flow velocity on vessel-side cooling.Source: §3.2 · p. 6

4 Results

The heat-sink effect was strongest when the tumor directly contacted the aorta, smaller with a 4 mm gap, and weak at 8.5 mm. Blood presence reduced vessel-side heating, while added flow velocity had little effect when a tissue gap was present.

4.1 Distance-Dependent Comparison

  • At D = 0 mm, Ctumor 60 coverage fell from 66.44% without blood to 52.59% with stagnant blood and 49.71% with normal flow.The blood-present contours were compressed near the vessel side.Source: Figure 3 · p. 7, §4.1 · p. 7
  • At D = 4 mm, Ctumor 60 coverage was 66.48% without blood, 65.25% with stagnant blood, and 65.27% with normal flow.The two blood-present values differed by only 0.02 percentage points.Source: §4.1 · p. 7
  • At D = 8.5 mm, Ctumor 60 remained between 66.89% and 67.05%, with a maximum difference of 0.16 percentage points.The authors report that the heat-sink effect on tumor coverage was weak at this distance.Source: §4.1 · p. 8

4.2 Blood Flow Condition Comparison

  • Stagnant blood reduced vessel-side thermal expansion, showing that the blood domain absorbed heat even without flow.The largest change occurred in the direct-contact case.Source: §4.2 · p. 8
  • At D = 0 mm, Ctumor 60 fell from 52.59% with stagnant blood to 49.71% with normal flow; differences were only 0.02 and 0.12 percentage points at D = 4 mm and D = 8.5 mm.The added effect of blood velocity on tumor coverage was small when a tissue gap was present.Source: §4.2 · p. 8
  • Tumor–aorta distance governed whether vessel cooling reached the tumor, while blood-flow condition governed heat removal after interaction with the blood domain.The authors recommend considering tumor coverage, the vessel-side 60°C margin, and patient-specific vessel position in treatment planning.Source: §4.2 · p. 8

5 Conclusion

The ANSYS-based coupled thermal-fluid simulation found that proximity to the aorta and the presence of blood reduced tumor-vessel interface temperature and tumor coverage, most clearly in direct contact. The conclusion calls for considering vessel distance and blood flow in treatment planning.

5 Conclusion

  • The simulation found reduced interface temperature and tumor coverage with shorter tumor-to-aorta distance and a blood domain, most clearly at D = 0 mm.Coverage reduction was smaller at D = 4 mm and nearly absent at D = 8.5 mm.Source: §5 · p. 8
  • The findings support considering tumor-to-aorta distance and blood flow in microwave ablation treatment planning.The authors also suggest simulation can help predict vessel-side heat loss before the procedure.Source: §5 · pp. 8–9
  • Future work could incorporate varied or randomly generated tumor geometries and machine-learning techniques such as surrogate modeling.These approaches are proposed to support patient-specific ablation planning.Source: §5 · p. 9