Wearable ECG Quality Assessment: A Deep Learning and Ambulatory Context-Awareness Approach
Xiaopeng Mao, Marike Weisbjerg, Sadasivan Puthusserypady
TL;DR
Ambulatory ECG quality assessment matters because free-living recordings have particularly severe noise compared with in-hospital recordings. This paper trains a deep-learning signal-quality model on contextualized ambulatory ECG data and finds that normal-PQRST rhythms are more likely to be predicted correctly than pathological waveforms with different morphologies.
Problem
Free-living ambulatory ECG recordings have particularly severe noise compared with in-hospital recordings, making it important to prevent noise from obscuring the ECG.
Source: §I · p. 1Method
The study trains a deep-learning signal-quality model on contextualized ambulatory ECG data, evaluates it on public databases, and analyzes ECG noise alongside physical context.
Source: §I · p. 1Results
ECG rhythms with normal PQRST character were more likely to be predicted correctly, whereas pathological waveforms with different morphologies were more likely to lead to wrong predictions.
Source: §IV · p. 5Takeaways & Limitations
Identifying physical context in relation to ECG signal quality marks a small beginning in signal-context research.
Source: §IV · p. 5Because the clean and noisy ECGs were not reviewed by a cardiologist, the contextualized analysis results warrant caution.
Source: §IV · p. 5
Abstract
from arXiv · showhide
This paper presents and evaluates a Deep Learning-based (DL-based) Signal Quality Assessment (SQA) model to distinguish between clean and noisy ambulatory Electrocardiograms (ECG). The model is trained on Copenhagen Center for Health Technology-Contextualized Arrhythmia Database (CACHET-CADB), which, to the best of our knowledge, is the first ambulatory ECG database with both physical and patient-reported contextual data. The model shows stable performance on different databases such as MIT-databases and the latest PyhsioNet/Cinc Challenge 2021 databases. Subsequently, the paper demonstrates how complicated ECG noise can be investigated by the SQA model and the physical contextual data.
I. INTRODUCTION · II. METHODS · A. Datasets
The study evaluates a deep-learning ECG signal-quality model trained on contextualized ambulatory data across public databases and explores how physical context can inform ECG-noise analysis. It also defines the CACHET-CADB labels and the clean/noisy test task.
I. INTRODUCTION
- The study evaluates a deep-learning signal-quality model trained on CACHET-CADB across MIT-BIH-Arrhythmia and PhysioNet/Cinc Challenge 2021 databases.It also aims to assess the model’s strengths and weaknesses across databases.Source: §I · p. 1
- The study explores whether combining the model with physical ECG context can deepen insight into noise and help improve the recording environment.Source: §I · p. 1
- Free-living ambulatory ECG can have more severe noise than in-hospital recordings, making it important to prevent noise from obscuring the signal.Source: §I · p. 1
A. Datasets
- CACHET-CADB contains 259 days of continuous ambulatory ECG from 24 people: 21 with atrial fibrillation and three healthy individuals.Signals were sampled at 1,024 Hz from one precordial lead.Source: A. Datasets · p. 1
- Two cardiologists independently annotated 1,602 ten-second ECGs as normal sinus rhythm, atrial fibrillation, noise, or other.Source: A. Datasets · p. 1
- For signal-quality assessment, the labelled ECGs serve as the test set: normal sinus rhythm, atrial fibrillation, and other rhythms are “clean,” while noise is “noisy.”Source: A. Datasets · p. 1
B. Continuous Wavelet Transformation
The section describes converting 1D ECGs into 2D CWT scalograms, whose time-frequency representation enables image classification of clean versus noisy signals. It also specifies the frequency scaling used and the InceptionV3 classifier used in this work.
B. Continuous Wavelet Transformation
- CWT represents 1D ECGs as scalograms using the absolute CWT coefficients, enabling clean-versus-noisy image classification through time-frequency information.Plain ECG provides only a time-domain representation.Source: B. Continuous Wavelet Transformation · p. 1
- The CWT uses a time-scaled and time-shifted mother wavelet; its frequency uses 1,024 Hz ECG sampling, the wavelet’s central frequency, and a logarithmic scale.The logarithmic frequency scale better represents the scalogram.Source: B. Continuous Wavelet Transformation · pp. 1–2
- Figures 1 and 2 show scalograms of 10-second NSR and noise ECGs, respectively.Source: B. Continuous Wavelet Transformation · p. 2, Figure 1 · p. 2, Figure 2 · p. 2
- CWT has been used for binary and multiclass ECG SQA on MIT-BIH arrhythmia and PhysioNet/CinC Challenge databases,; this work uses InceptionV3 for classification.Huerta et al. used transfer learning to train a series of deep-learning models.Source: B. Continuous Wavelet Transformation · p. 2
C. Model architecture
The model uses frozen InceptionV3 feature extraction followed by dense layers, with dropout added to stabilize training.
C. Model architecture
- The 224 × 224 × 3 input passes through frozen InceptionV3 modules to produce 5 × 5 × 2048 features, which are flattened before dense layers; dropout in Dense layer 1 stabilizes training.Freezing all InceptionV3 parameters is used to prevent overfitting.Source: C. Model architecture · p. 2
D. Contextualized analysis
The contextualized analysis uses device-measured physical parameters alongside ECG quality assessment, with ECG preprocessing, segment generation, and model training also specified.
D. Contextualized analysis
- The analysis uses barometer- and accelerometer-measured physical parameters, including body positions and activity classes, for contextualization.The parameters are listed in Table III.Source: D. Contextualized analysis · p. 2, Table III · p. 2
- ECGs are bandpass-filtered at 0.5–50 Hz, smoothed, and resampled to 256 Hz; 20,000 clean and 20,000 noisy 10-second segments are generated, with the method adjusted for AFIB.Segment labels use decision rules combined with template matching from.Source: D. Contextualized analysis · p. 2
- Model training uses Adam, cross-entropy loss, Python and TensorFlow, and three nodes of four Tesla V100 32 GB GPUs with NVlink.The optimizer and loss are cited as and, respectively.Source: D. Contextualized analysis · p. 3
III. RESULTS · A. Performance on CACHET-CADB test set
On the CACHET-CADB test set, the paper presents the model’s performance as a confusion matrix and reports stable convergence of both loss curves under the optimized hyperparameters.
A. Performance on CACHET-CADB test set
- The model’s performance on the CACHET-CADB test set is presented as a confusion matrix.The supplied passages do not provide its cell values.Source: A. Performance on CACHET-CADB test set · p. 3, Table IV · p. 3
- Both loss curves converge, indicating stable learning with the optimized hyperparameters.The learning rate decays exponentially by 82% every 10 epochs, using a base of 0.18 and a step of 10.Source: A. Performance on CACHET-CADB test set · p. 3, Figure 5 · p. 3
B. Performance on MIT-BIH-NSR, MIT-BIH-Arrhythmia and MIT-BIH-Noise Stress Test Database
The model was evaluated on MIT-BIH-NSR, MIT-BIH-Arrhythmia, and MIT-BIH-Noise Stress Test data using Lead 2 ECG. It tended to fail on arrhythmia records with abnormal morphology, including paced beats combined with PVC.
B. Performance on MIT-BIH-NSR, MIT-BIH-Arrhythmia and MIT-BIH-Noise Stress Test Database
- Evaluation used three randomly chosen MIT-BIH-NSR records, the entire MIT-BIH-Arrhythmia database, and three pure noise signals, with Lead 2 ECG.The noise signals represented baseline wander, muscle artifact, and electrode motion artifact; Lead 2 was selected because it most clearly expresses the QRS complex.Source: B. Performance on MIT-BIH-NSR,… · p. 3
- The model tended to fail on arrhythmia records with abnormal morphology, such as paced beats combined with PVC.Source: B. Performance on MIT-BIH-NSR,… · p. 4
C. Performance on PhysioNet/CinC challenge 2021
Evaluations on the PhysioNet/CinC challenge 2021 databases show that ECG morphology deviations pose more difficulty for the model than changes in beat frequency.
C. Performance on PhysioNet/CinC challenge 2021
- Chapman-Shaoxing evaluation used 10-second clean ECGs from 10,646 heart patients, with model performance categorized by rhythm.The database was sampled at 500 Hz; 10,247 ECGs were available from the challenge.Source: C. Performance on PhysioNet/CinC… · p. 4, Table VI · p. 4
- On Ningbo, the largest 2021 challenge database, evaluation covered all ECG types using manually identified annotations across 40,258 10-second ECGs from different patients.34,905 ECGs were available from the challenge, and recordings were sampled at 500 Hz.Source: C. Performance on PhysioNet/CinC… · p. 4
- The model is more error-prone when ECG morphology departs substantially from normal sinus rhythm, while sinus bradycardia or tachycardia does not severely affect performance.Source: C. Performance on PhysioNet/CinC… · p. 5
D. Contextualized analysis
Context histograms show noisy ECGs concentrated during upright sitting/standing and unknown context combinations, while upright sitting/standing is also most frequent for clean ECGs. The analysis suggests investigating unknown contexts by tracing their time periods and asking patients what was happening.
D. Contextualized analysis
- Noisy ECGs were mostly recorded during upright sitting/standing and combinations with unknown context, while upright sitting/standing was also the most frequent context for clean ECGs.Source: D. Contextualized analysis · p. 5, Figure 6 · p. 5, Figure 7 · p. 5
- Avoiding periods with unknown body position and activity class might improve overall signal quality.Source: D. Contextualized analysis · p. 5
- Unknown context parameters can be investigated by backtracking their time periods and asking patients about the underlying events.Source: D. Contextualized analysis · p. 5
IV. DISCUSSION
The discussion finds that normal-PQRST ECG rhythms were more likely to be predicted correctly than pathological morphologies, and cautions that contextualized analyses rely on clean/noisy labels not reviewed by cardiologists.
- Sinus-wave predictions were particularly accurate, but predictions for abnormal atrial waves such as AFIB were difficult.Source: §IV · p. 5
- Abnormal ST segments, ST intervals, and cardiac blocks were difficult to predict; CACHET-CADB lacked patients with diseases inducing these rhythms, consistent with MIT-BIH-Arrhythmia results in Table V.Source: §IV · p. 5
- Normal-PQRST ECG rhythms were more likely to be predicted correctly, while pathological waveforms with different morphologies more often produced wrong predictions.Source: §IV · p. 5
- Because cardiologists did not review the clean/noisy ECGs, the contextualized analyses in Figures 6 and 7 warrant caution.Source: §IV · p. 5
- Identifying physical context in relation to ECG signal quality marks a small beginning in the broader area of signal-context research.Source: §IV · p. 5